Zhitong Finance App News, Paulette (300246.SZ) issued an announcement. The company recently received a notice from its wholly-owned subsidiary Tianjin Zhixin Hongda Medical Device Development Co., Ltd. (“Tianjin Zhixin”). The China Drug Administration (“State Drug Administration”) recently carried out a flight inspection of Tianjin Zhixin, and published the “Notice” (“Notice”) on the website of the State Drug Administration on November 19, 2026.
The “Notice” indicates that the inspectors in Tianjin have not actually carried out citrate ion concentration tests, and that there have been improper acts such as modifying, deleting, copying, and fabricating inspection records. The above situation does not meet the relevant requirements for quality control in the “Medical Device Production Quality Management Code”. The “Notice” requires territorial provincial drug supervision and administration departments to take control measures to suspend production in accordance with the provisions of Article 72 of the “Medical Device Supervision and Administration Regulations” and carry out supervision and sampling when necessary; seriously deal with suspected violations of the “Medical Device Supervision and Administration Regulations” and related regulations in accordance with the law; order enterprises to evaluate product safety risks, and recall related products in accordance with the provisions of the “Medical Device Recall Management Measures”; after the enterprise has completed the rectification of all defective projects, it must be reviewed and passed by the local provincial drug supervision and administration department before production can resume.
According to reports, Tianjin Zhixin is mainly engaged in the production and sale of hemodialysis concentrates and hemodialysis dry powders. The incident had little impact on the company's overall operating performance. I learned that the product involved in Tianjin Zhixin this time is hemodialysis A powder. Other products, hemodialysis A solution, B hemodialysis powder, and online dry powder products do not contain citric acid, and do not involve citrate ion testing programs. After the problem was discovered, Tianjin Zhixin dealt with the relevant responsible person in accordance with the company regulations and tested the citrate ion concentration on the hemodialysis A powder product. The test results met the requirements; at the same time, some batches of products were sent to the Tianjin Medical Device Quality Supervision and Inspection Center for testing by a third party, and the test results met the requirements. Tianjin Zhishin will apply for review by the Tianjin Drug Administration as soon as the rectification is completed to resume production as soon as possible.